Construction
Body: cast stainless steel (ANC3B 18/8SS) Endcap: stainless
steel (AISI 316) Faceplate: alumina ceramic
Range
of moisture
For bulk materials the sensor will measure up to the point
of saturation, typically 0-20% for construction materials.
Penetration
Approximately 75-100mm dependent on material.
Power
supply
+15Vdc to +30Vdc. 3 watts max.
Operating
temperature
0-60 degrees C. The sensor will not measure moisture in frozen
materials.
Refresh
rate
Outputs are updated approximately 25 times per second.
|
|
Analogue
output
Configurable 0-20mA or 4-20mA current loop source. For 0-10Vdc
output an external 500R load resistor should be connected
across the analogue output at the control system.
Discrete
inputs
The batch average start stop or moisture/temperature multiplexing
functions can be controlled via the digital inputs. Refer
to 'Connecting the Cable' section of the manual for more information.
Digital
Communications
Opto-isolated RS485 2-wire port. Refer to your distributor
for read/write access to sensor parameters and values.
Grounding
The sensor body is connected to the cable shield. Ensure equipotential
bonding of all exposed metalwork. In areas of high lightning
risk correct and adequate protection should be used.
|