USING
THE HYDROPROBE II IN 'COMPATIBILITY MODE'
Compatibility
mode' can be used when replacing a HydroProbe in an existing application
with a HydroProbe II .
The
existing probe will already be connected in place using an extension
cable terminated in a 6-pin Bulgin connector at one end and a screw
terminal block at the other (for connection to a HydroView or HydroControl
IV system).
It
is possible to make use of the existing field wiring and HydroView
unit by fitting the HydroProbe II with a 4m 'Compatibility cable'.
The
required connections are shown opposite for reference. Note that
these are generally as given for the HydroView to HydroProbe connection
details with the following exceptions:
1.
-15V is no longer required, but may be left in place.
2.
The cable screen should be connected to ground at the HydroView
end of the cable.
N.B.
The HydroProbe II sensor may be purchased in 'Compatibility mode'
or can be configured for use in 'Compatibility mode' using the Hydro-Com
Diagnostics software.
|